Vertical autoclaves for your laboratory
Vertical autoclaves with a chamber volume of 80-150 litre
The Laboratory- Vertical- Standing- Autoclaves, short LVSA, are characterized by their low loading height and easy operation.
Our autoclaves are available in four different sizes:
Dimensions:
Model | LVSA 40/60 | LVSA 40/75 | LVSA 50/60 | LVSA 50/70 |
---|---|---|---|---|
Effective space in mm | (Ø)400 x (H)625 | (Ø)400 x (H)775 | (Ø)500x (H)655 | (Ø)500x (H)755 |
Chamber volume in liters | 80 | 96 | 133 | 153 |
Loading height (mm) | 760 | 900 | 800 | 900 |
External diameter in mm | (W) 550 x (D) 850 | (W) 750 x (D) 950 | ||
Heating capacity (kW) | 6 | 9 | ||
Electrical connection | 380 – 400V, 50/60 Hz, 16 A |
The following equipment variants are available:
Basic LabRecommended for the sterilization of simple materials or liquids without removal dry. | Integrated steam generatorSteriTouch control |
Dry LabRecommended for the sterilization of non-infectious waste bags and materials which require intensive subsequent drying. | Integrated steam generatorVacuum pump setSteriTouch control |
Quick Lab Basic Recommended for the sterilization of large quantities of liquids. | Integrated steam generatorRapid water recoolingsystem with air-circulating fanSteriTouch control |
Quick LabRecommended for the sterilization of non-infectious waste bags, large quantities of liquids and materials which require intensive subsequent drying. | Integrated steam generatorVacuum pump setRapid water recoolingsystem with air-circulating fanSteriTouch control |
Quick Lab AdvancedRecommended for the sterilization of non-infectious waste bags, large quantities of liquids and materials which require intensive subsequent drying – this also applies to liquids in closed containers. | Integrated steam generatorVacuum pump setRapid water recoolingsystem with air-circulating fanSteam/air mixture method (SAMM)SteriTouch control |
Safe LabRecommended for operation in the level S2 lab. Recommended for the sterilization of infectious waste bags and materials which require intensive subsequent drying. | Integrated steam generatorVacuum pump setExhaust air filtration EAFSteriTouch control |
Complete LabFully equipped laboratory autoclave for operation in the level S2 lab. Recommended for the sterilization of infectious waste bags, large quantities of liquids and materials which require intensive subsequent drying. | Integrated steam generatorVacuum pump setExhaust air filtration EAFRapid waterrecoolingsystem with air-circulating fanSteriTouch control |
Zirbus vertical autoclave LVSA 50/60
Loading capacity
(Specifications partly only possible without loading baskets!)
Medium-Flasks (Schott) | LVSA 40/60 | LVSA 40/75 | LVSA 50/60 | LVSA 50/70 |
---|---|---|---|---|
500ml | 2×13 | 3×13 | 2×23 | 3×23 |
1000ml | 2×9 | 3×9 | 2×16 | 3×16 |
Erlenmeyer flasks | LVSA 40/60 | LVSA 40/75 | LVSA 50/60 | LVSA 50/70 |
---|---|---|---|---|
500ml | 2×8 | 3×8 | 2×15 | 3×15 |
1000ml | 2×5 | 3×5 | 2×9 | 3×9 |
Description of the individual components
Steam generator– integrated in the housing of the autoclave. Ensures rapid heating and cooling compared to conventional heating systems.
Vacuum pump set– for mechanical removal of air and drying in vacuum. Possible methods include single-stage vacuum, fractionated vacuum, constant or pulsating drying under a vacuum.
Lifting and pivoting mechanism – makes loading and unloading of vertically loaded autoclaves easier. The load capacity is 30 kg.
Exhaust air filtration– is stipulated for the sterilization of pathogenic substances in S2/S3 laboratories, with inline condensate sterilization.
Touch display (SteriTouch control) – enables simple and intuitive operation at the autoclave. In combination with user-friendly software, the autoclave can be operated quickly but also highly variable and simple.
Steam-air mixture method – for the sterilization of tightly closed bottles and pressure-sensitive materials. An additional temperature sensor, air deflectors and a circulating air fan ensure precise temperature control and distribution.
Rapid water recooling system with air-circulating fan – allows a shortening of the recooling time by more than 80% compared to the self cooling during sterilization of liquids. This is due to an internal heat exchanger, also known as a “fast cooler”, through which water flows.
Accessories – We offer stainless steel sterilisation baskets, pails and containers which fit our various chamber designs.
Continue to the Laboratory-standing-autoclaves with up to 1.200 litres chamber volume.